PW-CUV40P UV Aging Test Chamber

Applicable standards: SAE J2020, PREN 1062-4, ISO4892-3, ASTM G53, EN534, ISO11507, Comply with ISO4892-3, ASTM G53 standards, etc. or set by yourself.

Function: automatic control of light irradiance, temperature, condensation and water spray.

Sample requirements: 

Prohibit the test or storage of flammable, explosive, and volatile substance samples, the test or storage of corrosive substance samples, the test or storage of biological samples, the test or storage of strong electromagnetic emission source samples.

Specification: 

Temperature range: Blackboard temperature (BPT): 40~75°C
Temperature fluctuation: ±1°C
Condensation method: nickel-chromium alloy electric water heating condensation system
Machine power: 2KW
Type of light source: U.S. Q-LAB company UVA-340
Power rate: 40W/one (U.S. Q-LAB, normal service life exceeds 5000 hours)
Wavelength range: UVA: 340nm UVA: 0.00~1.20 W/m
Internal material: stainless steel plate SUS304#
External material: stainless steel plate SUS304#
Air heating: nickel-chromium alloy heating, heating control method: non-contact SSR (solid state relay)
Water heating: nickel-chromium alloy heating, heating control method: non-contact SSR (solid state relay)
Number of light tubes: 8 pieces, UVA fluorescent uv light tubes
Man-machine touch screen: LCD display, easy to operate
Control system: adopt single-chip microcomputer control.
Sample rack: stainless steel metal, the distance between the sample surface and the center of the light pipe is 50±3mm, 24PCS.
Standard samples: 24 standard sample holders (75×300mm, that is, two 75×150mm standard samples).
Different test methods and procedures: automatic control, real-time monitoring and display
Sample rack temperature: BPT direct display and automatic control of sample rack temperature, and automatic calibration.
Condensation control: direct display and automatic control of condensation
Circulation mode: direct display and automatic control of light, condensation and spray Light irradiance: including irradiometer, the energy and time of the light irradiance can be set during the test.
Spray water: adjustable and displayable, automatic control, spray time can be set during the test.
Alarm function: automatic alarm function.
Water supply mode: automatic water supply.
Protection device: leakage protection, water shortage protection, heating overload protection, door safety protection, over temperature protection, automatic alarm for abnormal irradiance.
Dimensions: 1300×600×1500mm (W×H×D)